Daily cleaning

For N20 users

Note: Do not wash or wet the HumidX / HumidX Plus. When not in use, store in a clean and dry place.

  1. Clean and check the N20 connector by hand daily. Never wash it using a dishwasher or washing machine.
  2. Remove the HumidX before proceeding with cleaning.
  3. Wash the connector in warm water (~30°C or 86°F) and mild liquid detergent.
  4. Clean the connector using a soft bristle brush, with particular attention paid to the vent holes on the inside.
  5. Rinse well under running water, and then shake to remove excess water.
  6. Air dry out of direct sunlight, and check for dirt or dust before use.

For F20 users

  1. Clean and check the F20 connector by hand daily. Never wash it using a dishwasher or washing machine.
  2. Wash the connector in warm water (~30°C or 86°F) and mild liquid detergent.
  3. Clean the connector using a soft bristle brush, with particular attention paid to the vent holes on the outside.
  4. Rinse well under running water, and then shake to remove excess water.
  5. Air dry out of direct sunlight, and check for dirt or dust before use.

For P10 for AirMini users

  1. Take the mask apart into its main component parts, namely the nasal pillows, frame and mask tube.
  2. Remove the HumidX before proceeding with cleaning.
  3. Gently rub each part in warm water (~30°C or 86°F) and mild liquid detergent until clean, with attention paid to the pillows and its small openings.
  4. Rinse well under running water and allow to air dry out of direct sunlight.

Weekly cleaning

AirMini tubing

  1. Clean and check the AirMini tubing by hand weekly. Never wash it in a dishwasher or washing machine. 
  2. Separate tubing from ActiveAir vent connector. 
  3. Soak and wash the tubing in warm water (~30°C or 86°F) and mild liquid detergent.
  4. Rinse well inside and out under running water.
  5. Hang the tubing so that each end can drain, and leave it to air dry out of direct sunlight.
  6. Clean the AirMini by wiping down each side of its exterior carefully with a dry cloth.

Disassembly

For N20 users

  1. Disconnect the connector from the mask by pressing on its side buttons and pulling away. 
  2. Disconnect the vent module from the AirMini tubing by twisting them away from each other.
  3. Disconnect the vent module from the mask tube by twisting them away from each other. 
  4. If HumidX or HumidX Plus was used, remove it from the vent module.

For F20 users

  1. Disconnect the connector from the mask by pressing on its side buttons and pulling away. 
  2. Disconnect the F20 connector from the AirMini tubing by twisting them away from each other. 

For P10 users

  1. Disconnect the vent module from the AirMini tubing by twisting them away from each other. 
  2. Disconnect the vent module from the mask tube by twisting them away from each other. 
  3. If HumidX or HumidX Plus was used, remove it from the vent module.

AirMini

  1. Disconnect the AirMini tubing by pressing on the side of the grey cuff and pulling away. 
  2. Unplug the power supply unit from the power outlet and the machine.

Accessories best sellers

HumidX system

Exclusive to AirMini, HumidX and HumidX Plus are small heat and moisture exchangers (HME) that are designed to provide comfortable humidification.

Buy now

AirMini mount system

The ResMed mount system allows the AirMini device to be secured to various points such as a bedframe, bedside table, wall or airplane seat pocket.

Buy now

AirMini travel bag

Take your sleep therapy everywhere you go with the AirMini travel bag. Small and compact, it’ll fit the whole solution to keep everything together – so it’s always ready when you are.

Buy now

Inspection and replenishment

 

For F20 users

 

  1. Check the F20 connector daily for signs of damage or blockages caused by dirt and dust. 
  2. The connector should be replaced at least every six months or if there are signs of damage.

 

For N20 users

  1. Check the N20 connector daily for signs of damage or blockages caused by dirt and dust. 
  2. The connector should be replaced at least every six months or if there are signs of damage.

 

AirMini tubing

  1. Check the AirMini tubing daily for any holes, tears or cracks 
  2. The tubing should be replaced at least every six months or if there are signs of damage.

 

AirMini air filter

  1. Check the AirMini air filter for any holes, tears, cracks, or blockages caused by dirt or dust. 
  2. The air filter should be replaced at least every six months or if there are signs of damage. It cannot be washed. 
  3. To change the air filter, slide the air filter cover off and replace the old air filter with a new one. 
  4. Attach the air filter cover on again, and ensure it is fitted at all times to prevent water and dust from entering the machine.

 

HumidX and HumidX Plus

  1. Check your HumidX or HumidX Plus daily for signs of damage or blockages caused by dirt and dust. They cannot be washed. 
  2. HumidX and HumidX Plus must be replaced within 30 days after opening the packaging or if there are signs of damage.

Humidification

  1. To insert the optional HumidX / HumidX Plus, for N20 or AirFit P10 for AirMini users, open the connector by gently twisting. 
  2. Grip the HumidX / HumidX Plus at the edges, with blue / grey side facing down towards the tube, and place it inside. 
  3. Close the connector by twisting its ends back together until they lock into place. 
  4. HumidX and HumidX Plus must be discarded and replaced within 30 days after opening the HumidX / HumidX Plus packaging.

 

 

Accessories best sellers

HumidX system

Exclusive to AirMini, HumidX and HumidX Plus are small heat and moisture exchangers (HME) that are designed to provide comfortable humidification.

Buy now

AirMini mount system

The ResMed mount system allows the AirMini device to be secured to various points such as a bedframe, bedside table, wall or airplane seat pocket.

Buy now

AirMini travel bag

Take your sleep therapy everywhere you go with the AirMini travel bag. Small and compact, it’ll fit the whole solution to keep everything together – so it’s always ready when you are.

Buy now

References

  1. Netzel et al. APAP device technology and correlation with patient compliance. Somnologie - Schlafforschung und Schlafmedizin 2014; DOI: 10.1007/s11818-014-0662-0
  2. Isetta et al. Comparative assessment of several automatic CPAP devices responses: a bench test study. ERJ Open Res 2015;1:00031-2015
  3. Zhu et al. All APAPs are not equivalent for the treatment of sleep disordered breathing: a bench evaluation of eleven commercially available devices. J Clin Sleep Med 2015 11(7):725-34
  4. Isetta et al. Novel Approach to Simulate Sleep Apnea Patients for Evaluation Positive Pressure Therapy Devices. PLoS ONE 11(3): e0151530.doi:10.1371/journal.pone.0151530

Still have a question?

Contact us