Key features
Features | Stellar 150 |
---|---|
Modes |
iVAPS (intelligent Volume-Assured Pressure Support) with AutoEPAP ST (Spontaneous–Timed) PAC (Pressure Assisted Control) CPAP (Continuous Positive Airway Pressure) S (Spontaneous) T (Timed) |
Pressure range |
IPAP: 2–40 cm H2O EPAP: 2–25 cm H2O |
TiControl™ |
Ti Max 0.3–4 sec Ti Min 0.1–Ti Max |
Respiratory Rate | 5–60 bpm |
Rise Time | Min, 150–900 msec (approx.) |
Trigger and Cycle | Five sensitivity settings |
Adjustable alarms |
High Leak Low Minute Ventilation Low and High Pressure Low and High Respiratory Rate Apnoea Low and High FiO2 Low SpO2 Non-vented mask |
Standard fixed alarms |
Circuit disconnected Over pressure Blocked tube Internal battery empty |
Dimensions (LxWxH) | 230 x 170 x 120 mm |
Weight | 2.1 kg |
Air filter | 22 mm taper, compatible with ISO5356–1:2004 |
Air outlet | 22 mm taper, compatible with ISO5356–1:2004 |
Internal battery | Lithium-ion, 2 hours at EPAP 5, IPAP 15, 20 bpm |
Power supply |
AC 100–240V 50–60Hz, 2.2A AC 110V 400Hz, 2.2A DC 24V/3A |
IEC 60601–1 Classification | Class II (double insulation) Type BF Continuous operation |
Data solutions |
Airview ResScan EasyCare Tx software and Tx Link |