AirMini

Support

AirFit P10 for AirMini Usage Videos

 

Maintenance and inspection 

 

  1. Regular inspection of your mask and its components for signs of damage is important to maintain optimal performance of your mask.
  2. For a thorough inspection, you should first disassemble the mask as per instructions in your user guide and inspect each part individually.

    Check the mask frame and tube

  3. Examine both the inside and outside of the mask tube for any holes or tears. If damage is found, the frame will need to be replaced.
  4. Hold the frame and gently pull on each arm of your mask to ensure they are firmly attached and will provide good support.

    Checking the nasal pillows

  5. Examine your nasal pillows inside and out for general wear and tear, blockages, damage and deformed pillows.
  6. A light squeeze on the sides of deformed pillows can help bring them back to shape. Replace them if they are damaged or no longer fit well.

    Checking the headgear

  7. Hold the headgear in one hand and run your other hand along its length while stretching at various points to check for tears and its elasticity.
  8. If the headgear is torn, lost its elasticity or has damaged seams, you should replace it.

Accessories best sellers

 

 

HumidX system

Exclusive to AirMini, HumidX and HumidX Plus are small heat and moisture exchangers (HME) that are designed to provide comfortable humidification.

Read more

AirMini mount system

The ResMed mount system allows the AirMini device to be secured to various points such as a bedframe, bedside table, wall or airplane seat pocket.

Read more

AirMini Travel Bag

Take your sleep therapy wherever you go with the AirMini Travel Bag. Small and compact, it’ll fit the whole system to keep everything together – so it’s always ready when you are.

Read more

Disassembly

 

For N20 users

 

  1. Disconnect the connector from the mask by pressing on its side buttons and pulling away.
  2. Disconnect the vent module from the AirMini tubing by twisting them away from each other.
  3. Disconnect the vent module from the mask tube by twisting them away from each other.
  4. If HumidX or HumidX Plus was used, remove it from the vent module.

 

For F20 users

  1. Disconnect the connector from the mask by pressing on its side buttons and pulling away.
  2. Disconnect the F20 connector from the AirMini tubing by twisting them away from each other. 

 

For P10 users

  1. Disconnect the vent module from the AirMini tubing by twisting them away from each other.
  2. Disconnect the vent module from the mask tube by twisting them away from each other.
  3. If HumidX or HumidX Plus was used, remove it from the vent module.

 

AirMini

  1. Disconnect the AirMini tubing by pressing on the side of the grey cuff and pulling away.
  2. Unplug the power supply unit from the power outlet and the machine.

Daily cleaning

 

For N20 users
Note: Do not wash or wet the HumidX / HumidX Plus. When not in use, store in a clean and dry place.

 

  1. Clean and check the N20 connector by hand daily. Never wash it using a dishwasher or washing machine.
  2. Remove the HumidX before proceeding with cleaning.
  3. Wash the connector in warm water (~30°C or 86°F) and mild liquid detergent.
  4. Clean the connector using a soft bristle brush, with particular attention paid to the vent holes on the inside.
  5. Rinse well under running water, and then shake to remove excess water.
  6. Air dry out of direct sunlight, and check for dirt or dust before use.

 

For F20 users

 

  1. Clean and check the F20 connector by hand daily. Never wash it using a dishwasher or washing machine.
  2. Wash the connector in warm water (~30°C or 86°F) and mild liquid detergent.
  3. Clean the connector using a soft bristle brush, with particular attention paid to the vent holes on the outside.
  4. Rinse well under running water, and then shake to remove excess water.
  5. Air dry out of direct sunlight, and check for dirt or dust before use.

Weekly cleaning

 

AirMini tubing

 

  1. Clean and check the AirMini tubing by hand weekly. Never wash it in a dishwasher or washing machine.
  2. Separate tubing from AcitveAir vent connector.
  3. Soak and wash the tubing in warm water (~30°C or 86°F) and mild liquid detergent.
  4. Rinse well inside and out under running water.
  5. Hang the tubing so that each end can drain, and leave it to air dry out of direct sunlight.
  6. Clean the AirMini by wiping down each side of its exterior carefully with a dry cloth.

Accessories best sellers

 

 

HumidX system

Exclusive to AirMini, HumidX and HumidX Plus are small heat and moisture exchangers (HME) that are designed to provide comfortable humidification.

Read more

AirMini mount system

The ResMed mount system allows the AirMini device to be secured to various points such as a bedframe, bedside table, wall or airplane seat pocket.

Read more

AirMini Travel Bag

Take your sleep therapy wherever you go with the AirMini Travel Bag. Small and compact, it’ll fit the whole system to keep everything together – so it’s always ready when you are.

 

Read more

Inspection and replenishment

 

For F20 users

 

  1. Check the F20 connector daily for signs of damage or blockages caused by dirt and dust.
  2. The connector should be replaced at least every six months or if there are signs of damage.

 

For N20 users

 

  1. Check the N20 connector daily for signs of damage or blockages caused by dirt and dust.
  2. The connector should be replaced at least every six months or if there are signs of damage.

 

AirMini tubing

 

  1. Check the AirMini tubing daily for any holes, tears or cracks.
  2. The tubing should be replaced at least every six months or if there are signs of damage.

 

AirMini air filter

 

  1. Check the AirMini air filter for any holes, tears, cracks, or blockages caused by dirt or dust.
  2. The air filter should be replaced at least every six months or if there are signs of damage. It cannot be washed.
  3. To change the air filter, slide the air filter cover off and replace the old air filter with a new one.
  4. Attach the air filter cover on again, and ensure it is fitted at all times to prevent water and dust from entering the machine.

 

HumidX and HumidX Plus

 

  1. Check your HumidX or HumidX Plus daily for signs of damage or blockages caused by dirt and dust. They cannot be washed.
  2. HumidX and HumidX Plus must be replaced within 30 days after opening the packaging or if there are signs of damage.

Humidification

 

  1. To insert the optional HumidX / HumidX Plus, for N20 or AirFit P10 for AirMini users, open the connector by gently twisting.
  2. Grip the HumidX / HumidX Plus at the edges, with blue / grey side facing down towards the tube, and place it inside.
  3. Close the connector by twisting its ends back together until they lock into place.
  4. HumidX and HumidX Plus must be discarded and replaced within 30 days after opening the HumidX / HumidX Plus packaging.

Accessories best sellers

 

 

HumidX system

Exclusive to AirMini, HumidX and HumidX Plus are small heat and moisture exchangers (HME) that are designed to provide comfortable humidification.

Read more

AirMini mount system

The ResMed mount system allows the AirMini device to be secured to various points such as a bedframe, bedside table, wall or airplane seat pocket.

Read more

AirMini Travel Bag

Take your sleep therapy wherever you go with the AirMini Travel Bag. Small and compact, it’ll fit the whole system to keep everything together – so it’s always ready when you are.

Read more

*Combined responses of those expressing a preference in an independent patient survey in the USA, UK, Germany, France and Australia, from May 2015. Visit www.resmed.com/maskbrand